Low Latency MR Motion Phantom for MR-Linac Beam On/Off Latency Measurement
- Project
- 16016 STARLIT
- Type
- New standard
- Description
Provides capability to directly measure MR Linac Beam On/Off system latency without an oscilloscope when tracking a moving target as it moves in and out of a stationary pre-defined target volume.
- Contact
- Modus QA
- info@modusqa.com
- Technical features
Input(s):
- Beam On/Off Gating Signal
Main feature(s)
- Track moving dose target as it moves in and out of target volume
Output(s):
- Measure and report beam on and off latency times separately in ms with histograms
- Integration constraints
Requires beam on/off gating system signal as an input and real time MR motion tracking of a moving target provided by the Modus QA MRI 4D Motion Phantom and associated accessories.
- Targeted customer(s)
MR Linac Field Service/Installation Engineers and Clinical Physicist performing commissioning, acceptance testing, and AAPM TG recommended regular latency measurement QA.
- Conditions for reuse
Commercial distribution.